Features:
- Simultaneous sample processing, up to three times faster than comparable systems available in the market
- Up to 240 samples in 16 x 160 mm tubes or up to 80 samples in 32 x 115 mm tubes
- Superior handling of tablets, capsules, effervescent tablets, powder, pellets and liquids
- Brushless homogenizer motor for maintenance-free use
- Complete control via 21 CFR Part 11 compliant software with LIMS access
Sample Magazine and Sample Input:
- The unique Sotax designed sample magazine can run in a basic version
- With the extend version of the sample magazine it is possible to run up 240 samples (tube size 16 x 160 mm) or 80 (tube size 115 mm).
- The number of tests to be performed is not dependent on the number of dilution steps.
Solvent Input:
- With an optional rinse station the test tubes can be washed out with a selectable solvent to allow a quantitative transfer from your sample.
- The system covers a variety of solvent variants which with an 8-channel solvent selector allows greater flexibility for complex methods. Solvent introduction into the vessel is via a CP1-300 ceramic pump.
- The CTS has a built in temperature sensor, the software will automatically calculate the density of the solvent(s) to be dispensed.
Homogenizer:
- The unit is fully programmable up to a max speed of 15K rpm.
- The design feature of locating the bearings at the top of the tool ensures that at no time during the homogenization step is heat transferred to thesolvent during the process.
- Cleaning cycle requires a minimum amount of organic solvent for washing.
- Software allows programmable individual set up times for homogenization and wash cycle.
Sample Output and Sampling:
- Sample output is carried out via one CP1-35 ceramic head piston pump coupled to a standard filter station.
- The filter station gives the user the added flexibility of using industry standard filters from a range of manufacturers.
Test end and cleaning:
- Emptying and cleaning of the Vessels
- The Vessel is emptied and cleaned effectively and efficiently by nozzles located at angles in the cleaning station.
- Only approx. 25 ml of an organic solvent is required for washing, the vessel will be sprayed and dried with filtered air.
- Cleaning of the Homogenizers.
CTSoft Software:
- Fully 21 CFR Part 11 complaint
- High security through closed system with password access and different user rights assignable.
- Complete audit trail for access, test, method and hardware changes
- Electronic signatures
- Offline database available