Purifier Logic Class II Biosafety Cabinets
Purifier Cell Logic Class II, Type A2 Biosafety Cabinets are specially designed with cell researchers in mind, while also providing protection from hazardous particulates such as agents that require Biosafety Level 1,2 or 3 containment. These biohazard laminar flow cabinets are designed to discharge HEPA-filtered exhaust air directly into the laboratory, or into an exhaust system through the optional Canopy Connection. When canopy-ducted, these biological safety cabinets may be used for applications involving minute quantities of volatile toxic chemicals and tracer amounts of radionuclides as an adjunct to microbiological
  1. Features
  2. Specification

·                Patent-pending design

·                Interior-mounted, line-of-sight LCD information center with  “Filter Life Remaining” bar graph, status line for alarm conditions and alerts to warn when filter life diminishes to  20%, 10% and 0%

·                Filter monitoring system consisting of an electronically commutated motor (ECM) that delivers a precise volume of air as  required and automatically adjusts as filters load without relying on airflow sensors

·                Smart-Start™ System that allows the user to program start up  and shut down operations

·                Night-Smart™ System that idles the blower when the sash is  fully closed

·                Radiused type 304 stainless steel interior and removable

·                Epoxy-coated steel exterior

·                Contain-Air™ Negative Pressure Channel*

·                Class 5 conditions per ISO 14644-1 and 2 (formerly Class 100)

·                Supply and exhaust 99.99% efficient HEPA filters.

·                Fully-closing, clear 1/4" tempered safety glass sash with two sash handles

·                Bright, 90-100 footcandle, glare-free fluorescent lighting located outside the contaminated work area

·                Purifier Logic ClassII Biosafety Cabinet choose type A2 or type B2 with widths: 3 feet, 4 feet, 5 feet and 6 feet

Line-of-sight LCD information center

Fluororescen Light Level: 90-100 footcandle

Efficient HEPA filters: 99.99% or options 99.999%

Air recirculation:

+ Type A2: 70%

+ Type B2: 0%

UV Lamp 254nm

Sash opening viewing height: 21.7” (54cm)

Nominal inflow velocity and nominal downflow velocity: 105 fpm (0.5 m/s); 55 fpm (0.3 m/s)

Noise level: 62 dBA

Overall dimension and Weight:

3 feet: WxDxH (42”x31”x61”); 220 kg

4 feet: WxDxH (54”x31”x61”); 265 kg

5 feet: WxDxH (66”x31”x61”); 324 kg

6 feet: WxDxH (78”x31”x61”); 431 kg

Electrical Requirement: 220V/ 50-60Hz